Comparison of Software and Physical Grids for digital Mammography

https://doi.org/10.52560/2713-0118-2025-6-68-79

Abstract

Objective. To evaluate the possibility of using the «SyGrid» digital image processing program for mammography (software grid) instead of the traditional movable grid.

Materials and methods. On the basis of the mammograph Mammo-4MT-Plus (JSC «MTL», Moscow), the «SyGrid» software grid was compared with a standard movable grid. At the laboratory stage, CDMAM and CIRS 010D phantoms were used. At the clinical stage, doses were compared in samples of 1063 studies with a physical grid and 1132 studies with a software grid. To compare the image quality, 200 random studies from each sample were independently blindly evaluated by three radiologists.

Results. At equal doses, the quality metrics calculated with CDMAM and CIRS 010D phantoms with a software grid were no lower than with a physical grid. Radiologists, on average, rated the quality of images with a software grid significantly higher than with a physical one. The average dose in the sample with a software grid was 22 % lower than in the sample with a physical grid.

Conclusion. The «SyGrid» software grid can be used instead of a movable physical grid for mammography.

About the Authors

Yu. E. Shunkov
MEDICAL TECHNOLOGIES Ltd
Russian Federation

Shunkov Yury Evgen'evich, Ph. D., Lead Research Engineer 

Moscow



S. A. Varlamov
MEDICAL TECHNOLOGIES Ltd

Varlamov Sergey Andreevich, Lead Laboratory Integrator Engineer

Moscow



A. V. Smirnova
MEDICAL TECHNOLOGIES Ltd

Smirnova Anastasia Vladimirovna, application specialist (radiologist)

Moscow



G. P. Korzhenkova
Federal State Budgetary Institution «National Medical Research Center of Oncology named after N.N. Blokhin» of the Ministry of Health of the Russian Federation

Korzhenkova Galina Petrovna, M. D. Med., Senior Researcher

Moscow



A. R. Dabagov
MEDICAL TECHNOLOGIES Ltd

Dabagov Anatoliy Rudol'fovich, Ph. D., President

Moscow



References

1. Vinogradov I. V., Shunkov Yu. E., Prokhorov A. V. Sposob podavlenija shuma na rentgenovskom izobrazhenii s pomoshh'ju iskusstvennoj nejronnoj seti/ Patent RU № 2822515. 2024. № 14. (In Russ.).

2. Mazurov A. I., Potraxov N. N. Vliyanie rasseyannogo rentgenovskogo izlucheniya na kachestvo izobrazheniya i metody` ego podavleniya. Medicinskaya texnika. 2014;5:12-15. (In Russ.).

3. Pozdnyakov D. V., Shunkov Yu. E., Kasiuk D. M., Dabagov A. R. Sposob udalenija sostavljajushhej ot fonovogo rassejanija s rentgenograficheskogo izobrazhenija/ Patent RU № 2810896. 2023. № 1. (In Russ.).

4. Angelone F., Ponsiglione A. F., Grassi R., Amato F., Sansone M. A general framework for the assessment of scatter correction techniques in digital mammography. Biomedical Signal Processing and Control. 2024;89(6):1-12. https://doi.org/10.1016/j.bspc.2023.105802

5. Dance D. R., Skinner C. L., Young K. C., Beckett J. R., Kotre C. J. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys. Med. Biol. 2000;45(11):3225-3240. https://doi.org/10.1088/0031-9155/45/11/308

6. Fieselmann A., Fischer D., Hilal G., Dennerlein F., Mertelmeier T., Uhlenbrock D. Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality. Proceedings of SPIE Medical Imaging 2013: Physics of Medical Imaging. 2013;8668:86685Y. https://doi.org/10.1117/12.2007490

7. Ludewig E., Richter A., Frame M. Diagnostic imaging — Evaluating image quality using visual grading characteristic (VGC) analysis. Vet. Res. Commun. 2010; 34(5):473-479. https://doi.org/10.1007/s11259-010-9413-2

8. Marimon E., Marsden P. A., Nait-Charif H., Diaz O. A semi-imperical model for scatter field reduction in digital mammography. Phys. Med. Biol. 2021;66(4): 045001. https://doi.org/10.1088/1361-6560/abd231

9. Sharma R., Sharma S.D., Sarkar P. S., Datta D. Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System. J. Med. Phys. 2018;43(4):255-263.

10. Sayed M., Knapp K. M., Fulford J., Heales C., Alqahtani S. J. The principles and effectiveness of X-ray scatter correction software for diagnostic X-ray imaging: A scoping review. European J. Radiol. 2023; 158:110600. https://doi.org/10.1016/j.ejrad.2022.110600


Supplementary files

Review

For citations:


Shunkov Yu.E., Varlamov S.A., Smirnova A.V., Korzhenkova G.P., Dabagov A.R. Comparison of Software and Physical Grids for digital Mammography. Radiology - Practice. 2025;(6):68-79. (In Russ.) https://doi.org/10.52560/2713-0118-2025-6-68-79

Views: 11


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2713-0118 (Online)